Probable Causes
Suggested Actions
Troubleshooting and Test Functions

Leak Sensor Short

Leak Sensor Short
The leak sensor in the quaternary pump has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current
to change within defined limits. If the current increases above the upper
limit, the error message is generated.
• Defective leak sensor.
• Leak sensor incorrectly routed, being pinched by a metal component.
❏ Exchange the leak sensor.