Agilent Technologies 1100 Series Reference Manual page 225

Quaternary pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Table 37
Table 38
Table 39
Introduction to the Quaternary Pump
Setting the 8-bit Configuration Switch
Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the instrument must
be powered up again in order to store the values in the non-volatile memory.
Communication Settings for RS-232C Communication
Mode Select 1
2
RS-232
0
1
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.
Baudrate Settings
Switches
3
4
5
0
0
0
0
0
1
0
1
0
0
1
1
Data Bit Settings
Switch 6
Data Word Size
0
7 Bit Communication
1
8 Bit Communication
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.
3
4
5
Baudrate
Baud Rate
Switches
3
9600 (default)
1
1200
1
2400
1
4800
1
6
7
8
Data Bits Parity
Baud Rate
4
5
0
0
9600
0
1
14400
1
0
19200
1
1
38400
225