Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual

Probe Calibration

To meet full user specifications, you need to adjust the supplied red (R), blue (B),
gray (G) and green (V) VPS410-II 10:1 voltage probes for optimal response.
To adjust the VPS410-II probes:
1. Connect the red probe from the red Input A BNC to the banana jack. See
Figure 5-19.
2. Press
3. Select Probe Type:
4. Press
5. Press
6.
Press
5-44
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Probe Cal
Figure 5-19. 10:1 Probe Calibration Connection
A
and then
F3
Voltage | Attenuation:
ENTER
.
F3
- PROBE A.
F1
and follow the instructions shown on the display.
- PROBE CAL
Probe Cal Ref.
to open the
menu.
Probe on A
10:1.
st8416_t.eps