Fluke ScopeMeter 190 II Series Service Manual page 78

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

15. Observe the trace and check if the Test Tool triggers on the negative pulse
before the marker.
16. Select the Test Tool setup and press
menu.
17. Choose
VIDEO on A...
Polarity: NEGATIVE | PAL
18. Set the calibrator video trigger output signal to -100 %
19. Use
to select line number 310 (PAL, PALplus, or SECAM) or
262 (NTSC).
20. Set the calibrator format and marker line number to:
PAL 310 (odd) for PAL and PALplus
SECAM 310 (odd) for SECAM
NTSC 262 odd for NTSC
21. Observe the trace and check if the Test Tool triggers on the positive pulse
before the marker.
Figure 4-12. SC600 Marker Pulse
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
F4
to open the Trigger Options
and from the opened menu choose:
(or
or
or
NTSC
PALplus
Performance Verification
Scope Input A, B, C, D Tests
)
SECAM
video-sc600.bmp
4
4-33