Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
METER
A
2 CHANN. + METER
3. Set the 5502A to the first test point in Table 5-20. Use the 5502A "COMP 2
wire" mode for the calibration adjustments up to and including 100 kΩ. For
the higher values, the 5502A will turn off the "COMP 2 wire" mode.
4. Set the 5502A to operate (OPR).
5. Press
6. Wait until the display shows the calibration status
7. Press
calibration point, and start the calibration. Continue through all calibration
points.
8. When you are finished, set the 5502A to Standby.
9. Continue at the Save Calibration Data and Exit section.
CL 0910
CL 0911
CL 0912
CL 0913
CL 0914
CL 0915
5-42
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
USE STACKABLE TEST LEADS
AS SUPPLIED WITH CALIBRATOR
B
NORMAL
FLUKE 5502A CALIBRATOR
Figure 5-18. Four-wire Ohms Calibration Connections
F3
to start the calibration.
F2
to select the next calibration step, set the 5502A to the next
Table 5-20. Ohm Gain Calibration Points
Cal Step
SCOPE
.
:READY
UUT input Value (5500 NORMAL)
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
Perf-Ver-Z2.eps