4. Set the 5502A to operate (OPR).
F3
5. Press
to start the calibration.
6. Wait until the display shows calibration status
F2
7. Press
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points of Table 5-6.
8. Set the 5502A to Standby, and continue at the Multimeter (DMM) Meter Zero
section.
Table 5-6. Input ABCD Gain Calibration Points
Cal step
CL 0799
CL 0800
CL 0801
CL 0802
CL 0803
CL 0804
CL 0805
CL 0806
CL 0807
CL 0808
CL 0809
CL 0810
CL 0811
CL 0812
CL 0813
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Final Calibration For v11.10 and later
.
:READY
UUT input value (5502A NORMAL, dc output)
5 mV
12.5 mV
25 mV
50 mV
125 mV
250 mV
500 mV
1.25 V
2.5 V
5 V
12.5 V
25 V
50 V (set 5502A to OPR!)
125 V
250 V
5
Calibration Adjustment
5-23