Table 5-2. Input B LF-HF Gain Calibration Points
Cal step
UUT input signal
CL 0674
0.5 Vpp square wave, 1 kHz
CL 0435
100 mVpp square wave, 1 kHz
CL 0530
2.5 Vpp square wave, 1 kHz
CL 0582
2.5 Vpp sine wave, 50 kHz
CL 0583
2.5 Vpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
CL 0482
500 mVpp sine wave, 50 kHz
CL 0483
0.5 Vpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
CL 0462
100 mVpp sine wave, 50 kHz
CL 0463
100 mVpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
Fluke 190-202/204:
Fluke 190-102/104:
Fluke 190-062:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SCOPE edge, 0.5 Vpp, 1 kHz
SCOPE edge, 100 mVpp, 1 kHz
SCOPE edge, 2.5 Vpp, 1 kHz
SCOPE levsine, 2.5 Vpp, 50 kHz
SCOPE levsine, 2.5 Vpp,
501 MHz
501 MHz
381 MHz
381 MHz
151 MHz
151 MHz
111 MHz
111 MHz
SCOPE levsine, 500 mVpp, 50 kHz
SCOPE levsine, 0.5 Vpp,
501 MHz
501 MHz
381 MHz
381 MHz
151 MHz
151 MHz
111 MHz
111 MHz
221 MHz
221 MHz
141 MHz
141 MHz
91 MHz
91 MHz
SCOPE levsine, 100 mVpp, 50 kHz
SCOPE levsine, 100 mVpp,
501 MHz
501 MHz
381 MHz
381 MHz
151 MHz
151 MHz
111 MHz
111 MHz
221 MHz
221 MHz
141 MHz
141 MHz
91 MHz
91 MHz
Calibration Adjustment