Fluke ScopeMeter 190 II Series Service Manual page 106

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

F2
7. Press
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points of Table 5-5.
8. Set the 5502A to Standby, and continue at the Input Pos ABCD (AB)
Calibration section.
Table 5-5. Input AB Gain and Position Calibration Points
Cal step
UUT input value (5502A SCOPE, MODE volt, SCOPE Z 1 MΩ, 1 kHz)
CL 0504
500 mV, 1 kHz, MODE VOLT
CL 0624
Open inputs, Calibrator STBY
CL 0673
50 V, 1 kHz, MODE volt
CL 0660
300 mV
CL 0604
500 mV
CL 0637
CL 0504
500 mV
CL 0624
CL 0599
CL 0600
CL 0601
CL 0602
100 mV
CL 0622
CL 0603
250 mV
CL 0662
CL 0605
CL 0606
CL 0607
CL 0664
CL 0608
CL 0609
CL 0610
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
none
(5500 standby)
none
(5500 standby)
10 mV
25 mV
50 mV
none
(5500 standby)
2 V
1 V
2.5 V
5 V
20 V
10 V
25 V
50 V
(set 5502A to OPR!)
Calibration Adjustment
Final Calibration For v11.10 and later
5
5-21