Table 4-3. Volts DC Measurement Verification Points
Range
5502A Output V dc
+6.0 mV
2 mV/div
-6.0 mV
+15.0 mV
5 mV/div
-15.0 mV
+30.0 mV
10 mV/div
-30.0 mV
+60.0 mV
20 mV/div
-60.0 mV
+150 mV
50 mV/div
-150 mV
+300 mV
100 mV/div
-300 mV
+600 mV
200 mV/div
-600 mV
+1.50 V
500 mV/div
-1.50 V
+3.00 V
1 V/div
-3.00 V
+6.00 V
2 V/div
-6.00 V
+15.0 V
5 V/div
-15.0 V
+30.0 V
10 V/div
-30.0 V
+60.0 V
20 V/div
-60.0 V
+150 V
50 V/div
-150 V
+300 V
100 V/div
-300 V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2 Channel Input Reading
+4.9 to +7.1
-7.1 to -4.9
+14.2 to +15.8
-15.8 to -14.2
+28.9 to +31.1
-31.1 to -28.9
+58.5 to +61.5
-61.5 to -58.5
+142 to +158
-158 to -142
+289 to +311
-311 to -289
+585 to +615
-615 to -585
+1.42 to +1.58
-1.58 to -1.42
+2.89 to +3.11
-3.11 to -2.89
+5.85 to +6.15
-6.15 to -5.85
+14.2 to +15.8
-15.8 to -14.2
+28.9 to +31.1
-31.1 to -28.9
+58.5 to +61.5
-61.5 to -58.5
+142 to +158
-158 to -142
+289 to +311
-311 to -289
Performance Verification
Scope Input A, B, C, D Tests
4 Channel Input Reading
+4.8 to +7.2
-7.2 to -4.8
+13.9 to +16.1
-16.1 to -13.9
+28.5 to +31.5
-31.5 to –28.5
57.6 to 62.4
-62.4 to -57.6
+139 to +161
-161 to -139
+285 to +315
-315 to -285
+576 to +624
-624 to -576
+1.39 to +1.61
-1.61 to -1.39
+2.85 to +3.15
-3.15 to -2.85
+5.76 to +6.24
-6.24 to -5.76
+13.9 to +16.1
-16.1 to -13.9
+28.5 to +31.5
-31.5 to –28.5
57.6 to 62.4
-62.4 to -57.6
+139 to +161
-161 to -139
+285 to +315
-315 to -285