Instrument Communication - Hach 2100AN Instrument Manual

Laboratory turbidimeter
Hide thumbs Also See for 2100AN:
Table of Contents

Advertisement

SECTION 6,
continued

6.3 Instrument Communication

Note: A specified cable or
equivalent is mandatory for EC
compliance (a shielded cable
assembly is required).
Figure 19
Typical 2100AN to Serial Printer Cable
Figure 20
Printer Format Example
6. Press
.
ENTER
7. Press
to exit the Setup mode.
SETUP
The instrument can communicate with either a serial printer or a serial
communication port on a computer (see Figure 19 and Figure 20). If the RS232
feature is used for a serial printer, a printer cable assembly terminated with a
standard 25-pin D connector is an optional accessory (refer to SECTION 13 on
page 99). Using a serial-to-parallel converter, the data string transmitted from the
2100AN Turbidimeter prints on any Epson compatible parallel printer that is used
with IBM compatible applications.
Data is transmitted to the printer as a 39-character string plus the line feed and
carriage return.
HACH 2100AN V1.0
01/01/94 12:00:00:00
01/01/94 12:00:00:00
0.009 NTU SAMPLE#00
CALIBRATION DATA
UNITS: NTU
DATE: 01/01/94 12:00:00:00
STANDARDS:
00 0.0434
01 20.000
02 200.00
03 1000.0
04 4000.0
05 7500.0
COEFFECIENTS:
A0=602.77400
B0=0.0022002
B1=0.0006147
C0=0.0022703
C1=0.0006442
C2=-1.000184
D0=0.0022970
D1=0.0006096
D2=-0.000034
D3=-0.000126
69