Report output

N O T I C E
The data output (printer or PC) needs to be selected first, so that the Type of
report menu is available (refer to
Select the data output
Different report output types can be selected when a printer or a PC is
connected.
1. From the main menu use the
2. Use the
or
to select Type of report and confirm.
3. When a printer or a computer and Terminal is connected, use the
or
to select
Option
Description
Reduced Select Several or One sample as an output format
Standard Select Several or One sample as an output format. Select
several: Users: The user name appears on the printed report
(17 characters). Header:The company name can be added as a
header (40 characters) and appears on the print report. Identify
sensor:The sensor model and the sensor serial number can be
added and appears on the print report.
GLP
Select Several or One sample as an output format. Select
several: Users: The user name appears on the printed report
(17 characters). Header:The company name can be added as a
header (40 characters) and appears on the print report. Identify
sensor:The sensor model and the sensor serial number can be
added and appears on the print report.
4. When a computer is connected and LabCom Easy (refer to
the data output
on page 11 for more information) is selected, use the
or
to select
Option
Description
Users
The user name appears on the printed report
(17 characters).
Identify sensor The sensor model and the sensor serial number can be
added and appears on the print report.
on page 11).
or
to select SYSTEM. Confirm.
Select

Maintenance

W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance or service.
If the internal components must be cleaned or repaired, contact the
manufacturer.
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks
described in this section of the manual.

Clean the instrument

N O T I C E
Never use cleaning agents such as turpentine, acetone or similar products to
clean the instrument including the display and accessories.
Clean the exterior of the instrument with a moist cloth and a mild soap
solution.
Clean the probe
Clean the probe as needed. Refer to
more information about cleaning. Refer to the probe documentation for
information about the probe maintenance.
Use the cleaning agents listed in
probe.
Table 1 Cleaning agents for the pH probe
Contamination
Proteins
Grease, oils, fats
Limescale
Troubleshooting
on page 16 for
Table 1
for contaminations on the pH
Cleaning agent
Pepsin cleaning solution
Electrode cleaning solution
0.1 N HCl solution
English 13