Weighing Functions
Procedures:
Check all parameters and change if necessary.
o
Send the DS3 to the scale (Not possible with the wire bridge on terminal P-PR,
o
the data is not accepted)
Perform the scale adjustment
o
Receive DS3 from the scale
o
Name
Adjustment digits for zero
point
Adjustment digits for
Adjustment weight 1
Adjustment digits for
adjustment weight 2
Adjustment weight 1
Adjustment weight 2
Characteristic value
range
Measurement time
Limit frequency for low
pass filter fg
Filter depth - Average
value filter
Scale Name
Min. weighing range
Weighing range
Numeral step
Standstill Range
Standstill Time
5-20
SIWAREX MS
Type
Length
Default
WORD
2
5461
WORD
2
60074
WORD
2
INT
2
2000
INT
2
BYTE
1
BYTE
1
BYTE
1
BYTE
1
STRING[10]
12
"SIWARE
X MS"
INT
2
INT
2
2000
INT
2
INT
2
TIME
4
1000
Range of Values / Meaning
Scale zero point (JD0) [ 0 to: 2
Other definitions not permitted.
For description see chapter
Adjustment digits for adjustment weight 1 (JD1) [ 0 to: 2
Other definitions not permitted.
For description see chapter
0
Adjustment digits for adjustment weight 2 (JD2) [ 0 to: 2
Other definitions not permitted.
For description see chapter
Adjustment weight 1 (JG1)
For description see chapter
0
Adjustment weight 2 (JG2)
For description see chapter
2
1: Characteristic value range 1mV/V
2: Characteristic value range 2mV/V
4: Characteristic value range 4mV/V
Other definitions not permitted.
For description see chapter
0
Bit 0 Measurement time
0 = the measurement time is 20 ms (Sample rate 50 Hz)
1 = the measurement time is 33 1/3 ms (Sample rate 30 Hz)
For description see chapter
Bit 2 to 7 not used
4
3: fg = 5Hz
4: fg = 2Hz
5: fg = 1Hz
6: fg = 0.5Hz
7: fg = 0.2Hz
8: fg = 0.1Hz
9: fg = 0.05Hz
Other definitions not permitted.
For description see chapter
15
Filter depth of average value filter [0 .. 255]
0 and 1: Average value filter deactivated
>1: Filter depth
For description see chapter
Customer specific scale designation
For description see chapter
20
Minimum Weight for Weighing Range
For description see chapter
Maximum Weight for Weighing Range
For description see chapter
1
Numeral step (1, 2, 5, 10, 20)
For description see chapter
10
Standstill range in weight unit
For description see chapter
Standstill time in ms
For description see chapter
16
]
5.2.1
5.2.1
5.2.1
5.2.1
5.2.1
5.2.2
5.2.3
5.2.4
5.2.6
5.2.7
5.2.8
5.2.9
5.2.10
5.2.12
5.2.11
16
]
16
]