Specifications; Chapter 1 Specifications - Emerson Rosemount 370XA Quick Start Manual

Enclosure natural gas chromatograph
Hide thumbs Also See for Rosemount 370XA:
Table of Contents

Advertisement

Quick Start Guide
9-00200-954
1

Specifications

Minimum requirements
These are the minimum requirements for a typical installation. Please reference the Rosemount 370XA
Reference Manual
for more details or call the factory for additional support.
24 Vdc (21 Vdc to 30 Vdc)
55 Watts startup, < 25 Watts steady state
Environmental temperature
-4 to 140 ºF (-20 ºC to 60 ºC)
Heater standard power
120 Vac or 230 Vac, 300 W
Junction box protection rating
NEMA 4X
Carrier gas
Must be regulated to 90 psig (6.2 BarG)
Zero-grade helium
Actuation gas
Must be regulated to 90 psig (6.2 BarG)
Helium
Nitrogen
Clean, dry air
Rosemount 370XA Enclosure