4. Configure the Ethernet port.
Figure 10-4: TCP/IP Settings Screen
a) From the Main Menu, use the arrow keys to navigate to the Application menu and select the
TCP/IP Settings option.
b) Use the arrow keys to navigate through the various settings and press the Select/Edit key to edit
the appropriate values.
The settings must match the settings of the host device communicating to the Rosemount
370XA on that port.
c) When you have finished making changes, press
5. Enter the calibration gas values.
a) From the Main Menu, navigate to the Application menu and select Calibration Gas Info.
Figure 10-5: Calibration Concentration Screen
b) Press Select/Edit to enter the calibration gas concentration values for each component.
Note
The Methane value is calculated automatically. You can use this value as a check against the
value on the certificate to ensure all the values have been entered correctly.
Rosemount 370XA Enclosure
Configure and calibrate the Rosemount 370XA gas chromatograph
to save changes and close the screen.
November 2019
31