Emerson Rosemount 370XA Quick Start Manual page 33

Enclosure natural gas chromatograph
Hide thumbs Also See for Rosemount 370XA:
Table of Contents

Advertisement

Quick Start Guide
9-00200-954
This should take approximately two hours from when power is applied.
7. Clear alarms.
Figure 10-9: Current Alarms Screen
a) From the Main Menu, navigate to the View menu and select Current Alarms.
b) Press 2 to acknowledge and clear all alarms.
c) Press
to return to the Main Menu.
8. Purge calibration gas.
Figure 10-10: Select Cal Gas for a Single Stream Analysis
a) From the Main Menu, navigate to the GC Control menu and select Single Stream.
b) Select 4-Cal stream and the Purge Stream for 60 seconds option.
c) Let the GC run for at least 30 minutes.
Rosemount 370XA Enclosure
Configure and calibrate the Rosemount 370XA gas chromatograph
November 2019
33