Emerson Rosemount 370XA Quick Start Manual page 12

Enclosure natural gas chromatograph
Hide thumbs Also See for Rosemount 370XA:
Table of Contents

Advertisement

Mounting the enclosure
November 2019
Note
Minimum edge distance 18 in. (457.2 mm) (edge of concrete to edge of enclosure of all four sides).
2. Use a forklift or slings to place the enclosure on the mounting surface.
See
Figure 3-2
for proper positioning of forklift tines or slings.
CAUTION
Lift the enclosure by the metal frame, not the glass fiber reinforced polyester box.
Figure 3-2: Proper Positioning of Forklift Tines or Slings
A. To lift, place forks or slings here.
3. Ensure that the enclosure's foot plate pre-drilled holes align with holes in the mounting surface.
4. Secure the enclosure to the mounting surface with the cement anchors.
12
Quick Start Guide
9-00200-954
Emerson.com/Rosemount