Configure and calibrate the
Rosemount 370XA gas
chromatograph
Prerequisites
As the GC warms up to operating temperature and purges the carrier gas through the system, configure the
GC's site-specific settings, such as the calibration gas values and communication settings.
Procedure
1. If the GC is not in Idle mode, do the following:
Figure 10-1: Home Screen
a) Press 3 on the keypad to go to the GC Control menu.
b) Press the Down arrow to highlight the Halt command.
c) Press
on the keypad and then follow the prompts.
The Login screen appears if you are not logged in.
d) Enter your username and password.
The default values for the Rosemount 370XA gas chromatograph are:
User: EMERSON
Password: (blank)
Rosemount 370XA Enclosure

Configure and calibrate the Rosemount 370XA gas chromatograph

November 2019
29