Specifications

3.1 

Measurement specifications

3.1.1 
Net O
Variable 0 - 10% to 0-50% (Xi electronics of 0-50% range)
3.1.2 
Accuracy in oxidizing condition
±0.75% of reading at 0.05% O
ranges and above.
3.1.3 
Lowest detectable limit
0.02% O
3.1.4 
Process temperature effect
Less than 0.05% O
3.1.5 
System speed of response to calibration gas
Initial response in less than 3 seconds; T90 in less than 8 seconds. Response to
process gas changes varies depending on process gas velocity and particulate
loading of the diffuser.
3.1.6 
Calibration validity
Presentation of calibration gases matches the bottle value to within 0.02% O
3.1.7 
Accuracy in reducing conditions (requires
stoichiometer feature)
±10% of reading or 0.1% O
3.1.8 
System response in reducing conditions (requires
stoichiometer feature)
Going from oxidizing to reducing - T90 in 120 seconds
Going from reducing to oxidizing - T90 in 30 seconds
Manual
range
2
2
2
from 100 to 700 °C (212 to 1292 °F)
2
, whichever is different size
2
, whichever is greater. ±1% of full scale for 0-10%
Specifications
January 2023
2
15