PREPARATION AND SAFETY CHECK
NOTE
Do not press the water bottle too forcefully into the EPK-
700. Mishandling may cause water to leak onto/into the
video processor.
5.
Insert the Air/Water connector into the holder on the water
bottle cap assembly until the endoscope is connected
NOTE
Always disconnect the water bottle before moving the
video processor into a position other than in normal use.
Always disconnect the water bottle before packing the
EPK-700 for shipment.
4-1-3

Connecting the endoscope

CAUTION
Always turn OFF the processor BEFORE connecting
an endoscope. Connecting the endoscope while the
processor is ON might damage the endoscope.
1.
Check to ensure the appropriate (light guide) adapter is
mounted onto the EPK-700. The adapter, AE-P1 is already
attached to the EPK-700.
NOTE
Connecting a video endoscope without a (light guide)
adaptor in place will reduce light output at the endoscope
distal end. Attempting to connect a fibrescope without a
(light guide) adapter and/or an appropriate (light guide)
sleeve in place may damage the fibrescope and the
EPK-700.
2.
Check to ensure that the (scope) locking lever is open.
3.
Connect the (scope) light guide into the processor.
4.
Connect the (scope) electrical connector into the processor.
5.
Close the (scope) locking lever
6.
Connect the water bottle Air/Water connector to the Air/Water
receptacle on the endoscope's umbilical connector.
7.
Connect the suction tube of the suction device to the suction
nipple on the umbilical connector of the endoscope.
16
EP K- 70 0
X
P E N T A
CO NT RO
L
E
BA LA NC
CO LO R
ES S
BR IGH TN
COP Y
+5
LA MP
AUT O
RED
SER VIC
E
MAN
BLU E
0
FRE EZE
AVE
CEN T
-5
XLU M
PO WE R
ON
OFF
ON
OFF