NOMENCLATURE, CONTROLS AND FUNCTIONS
3-2

Water bottle assembly, Model OS-H2

No. N ame
Bottle
Water Bottle Cap
Assembly
Air Pipe Stem
Air/Water Hose
Air/Water Connector
Air/Water-Drain Lever Must be set to upright (A/W) position for delivery of air and water.
Water Feeding Stem
NOTE
If the water bottle assembly has been mishandled, the water feeding tube inside the Air/Water hose may be disconnected at
the A/W connector for the endoscope. To test, remove the cap assembly and using a syringe, inject water into the water feeding
stem. If the water comes out of both the center hole of the A/W connector and the series of holes around the center hole, the
water feeding tube is disconnected. Use another water bottle assembly.
12
F unction
Holds sterile water for procedure. (use up to 2/3's full)
Must be firmly secured to bottle to prevent air leakage.
Inserts into video processor water bottle receptacle.
Contains (2) independent tubes - 1 for air, 1 for water.
Inserts into Air/Water socket of endoscope umbilical connector.
Channel for water to be displaced from bottle and into scope.