Installation; Injection Test - Panasonic MDF-DU702VH Service Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

INSTALLATION

1. Put the liquid CO
cylinder into the CO
2
each cylinder to the stand with chains.
2. Connect the valve on the CO
outlet of the cylinder with a flexible hose.
3. Close the valve on the CO
the cylinder valve. Then check for CO
valve, cylinder stand valve and the connecting area of the flexible
4. When no leakage is found, close the cylinder valve.
5. Cut the connecting pipe (total length; 2.5m) enclosed with the
backup cooling kit into an appropriate length. The cut end should
6. Remove the cap covering the connecting port on the backup
cooling kit. Place the flared nut, ring, and sleeve onto the end of
7. Insert the cut side of the connecting pipe all the way into the
connecting port on the backup cooling kit and tighten the flared
8. Connect the other side of the connecting pipe with the cylinder
valve (when 1 cylinder) or the cylinder stand valve (when more
than 1 cylinder). Tighten the flared nut all the way.
Note: Run the connecting pipe in a coil or U-bend formation to act
as a cushion (see Fig. 2).
9. Open the cylinder valve or cylinder stand valve (whichever
applicable) and check for leakage at the both ends of the
connecting pipe. The leakage can be checked by applying soapy
water to the connecting area as shown in Fig. 3.
10. Close all of the cylinder valve and also the cylinder stand valve.
11. Check the injection by following the procedure in the "Injection Test" on the next page.

INJECTION TEST

1. Run the freezer until the chamber temperature reaches the appropriate level.
2. With the power switch of the backup cooling kit turned off, open the cylinder valve and the cylinder stand valve.
3. Set the temperature of the backup cooling kit to a temperature 10℃ or more higher than the freezer chamber temperature.
Note:
The temperature setting range of the backup cooling kit is between -50 and -70℃. The consumption of liquid CO
gas is increased and the backup cooling time is decreased considerably if the set temperature is lower than -70℃.
4. Turn on the power switch of the backup cooling kit with the freezer door closed.
For the up-right freezer, check the indicator lamp (green ) on the door switch is ON.
5. Press the test switch for the backup cooling kit to check the liquid CO
With the back-up system set in this way, the liquid CO
freezing when the chamber temperature reaches the set temperature of the backup cooling kit.
6. If the set temperature of the backup cooling kit is below -60℃, turn off the power switch of the freezer.
7. Check the liquid CO
2
Note the backup cooling time is decrease if the liquid CO2 gas is injected continuously.
8. Ensure the injection of the liquid CO
●Refer to the "Operating Instructions" for an explanation of the switches on the freezer.
cylinder stand. Secure
2
cylinder stand with the CO
2
cylinder stand completely and open
2
leakage at the cylinder
2
gas is injected into the freezer chamber automatically when the chamber temperature rises.
gas automatically stops when the chamber temperature goes down.
2
2
gas is injected into the freezer chamber.
2
gas is injected into the freezer chamber automatically to ensure
2
-72-