Temperature Calibration Procedure - Panasonic MDF-DU702VH Service Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

(Temperature calibration procedure:chamber sensor)
Keep touching MENU KEY (5 sec)
Move to Menu display, touch Service #2
key.
Touch mark area.
Cal key is vialed within setting range, touch
Cal key.
NOTE:: Ambient Temp. sensor and condenser A&B sensor can calibrate in the same procedure.
-43.3
-43.3
Move to Service Code input display,
so input "335232"
Move to Service display, touch Temp.
Calibration key
Move to input value display.
Input measured value by thermometer
Touch TOP key, so return to Menu display.
Input value area return to zero.
-20-
-43.3