Internal Temperature Uniformity (Reference Data); Distribution Data - Panasonic MDF-DU702VH Service Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

Internal Temperature Uniformity (Reference Data)

MDF-DU502VH
<Conditions>
Ambient temperature: 23/30℃
Load: Unloaded
<Distribution data>
Temperature of the cycle in each area (SV=-80℃、air temperature)
230V,50Hz
Maximum
Minimum
-77.0
-79.6
Left front
-77.2
-80.1
Left back
-77.4
-80.0
Right front
-77.3
-79.9
Right back
-77.2
-79.5
Center
-79.8
-80.9
Left front
-80.3
-81.6
Left back
-79.7
-81.0
Right front
-80.2
-81.7
Right back
-80.4
-81.1
Center
-78.7
-80.6
Left front
-79.1
-81.2
Left back
-79.2
-81.2
Right front
-79.2
-81.5
Right back
-79.1
-81.3
Center
Center of Upper shelf
-79.3
-80.1
Center of Lower shelf
-80.5
-81.3
Average
-
-
230V,50Hz
Maximum
Minimum
-76.7
-79.4
Left front
-76.9
-79.9
Left back
-77.2
-79.8
Right front
-77.0
-79.7
Right back
-76.9
-79.3
Center
-79.8
-81.1
Left front
-80.4
-81.7
Left back
-79.8
-81.0
Right front
-80.3
-81.7
Right back
-80.5
-81.2
Center
-79.1
-81.2
Left front
-79.4
-81.5
Left back
-79.5
-81.4
Right front
-79.6
-81.8
Right back
-79.5
-81.6
Center
Center of Upper shelf
-79.3
-80.0
Center of Lower shelf
-80.7
-81.5
Average
-
-
Note:This data does not represent a guarantee of product performance.
<Amount of power consumption>
Amount of power consumption when driving at cycle
(SV=-80℃)
230V,50Hz
AT23℃
6.70
AT30℃
7.80
Note:This data does not represent a guarantee
of product performance.
Ambient temperature 23℃ 
220V,60Hz
Middle of
Middle of
Differential Maximum
Minimum
cycle
cycle
-78.3
±1.3
-76.9
-79.6
-78.3
-78.7
±1.5
-77.1
-80.0
-78.6
-78.7
±1.3
-77.4
-79.9
-78.7
-78.6
±1.3
-77.2
-79.8
-78.5
-78.4
±1.2
-77.1
-79.5
-78.3
-80.4
±0.5
-79.8
-80.8
-80.3
-81.0
±0.6
-80.3
-81.6
-81.0
-80.4
±0.6
-79.8
-80.8
-80.3
-81.0
±0.8
-80.2
-81.5
-80.9
-80.8
±0.4
-80.4
-81.0
-80.7
-79.7
±1.0
-78.7
-80.4
-79.6
-80.2
±1.1
-79.1
-81.1
-80.1
-80.2
±1.0
-79.2
-81.1
-80.2
-80.4
±1.2
-79.3
-81.4
-80.4
-80.2
±1.1
-79.1
-81.1
-80.1
-79.7
±0.4
-79.3
-80.0
-79.7
-80.9
±0.4
-80.5
-81.1
-80.8
-79.8
-
-
-
-79.8
Ambient temperature 30℃ 
220V,60Hz
Middle of
Middle of
Differential Maximum
Minimum
cycle
cycle
-78.1
±1.4
-76.8
-79.5
-78.2
-78.4
±1.5
-77.0
-80.0
-78.5
-78.5
±1.3
-77.2
-79.8
-78.5
-78.4
±1.4
-77.1
-79.8
-78.5
-78.1
±1.2
-77.1
-79.4
-78.3
-80.5
±0.6
-79.8
-81.1
-80.5
-81.1
±0.6
-80.3
-81.8
-81.1
-80.4
±0.6
-79.7
-80.9
-80.3
-81.0
±0.7
-80.3
-81.7
-81.0
-80.9
±0.4
-80.5
-81.2
-80.9
-80.2
±1.1
-79.2
-81.2
-80.2
-80.5
±1.1
-79.5
-81.5
-80.5
-80.5
±1.0
-79.6
-81.4
-80.5
-80.7
±1.1
-79.6
-81.8
-80.7
-80.6
±1.1
-79.5
-81.6
-80.6
-79.7
±0.3
-79.3
-80.0
-79.7
-81.1
±0.4
-80.8
-81.5
-81.2
-79.9
-
-
-
-79.9
Unit:kWh/day
220V,60Hz
6.80
7.90
<Distribution data>
Temperature of the cycle in each area (SV=-70℃、air temperature)
Unit:℃
Differential
±1.4
Left front
±1.5
Left back
±1.3
Right front
±1.3
Right back
±1.2
Center
±0.5
Left front
±0.6
Left back
±0.5
Right front
±0.6
Right back
±0.3
Center
±0.9
Left front
±1.0
Left back
±1.0
Right front
±1.1
Right back
±1.0
Center
±0.3
Center of Upper shelf
±0.3
Center of Lower shelf
-
Average
Unit:℃
Differential
±1.3
Left front
±1.5
Left back
±1.3
Right front
±1.4
Right back
±1.2
Center
±0.6
Left front
±0.8
Left back
±0.6
Right front
±0.7
Right back
±0.4
Center
±1.0
Left front
±1.0
Left back
±0.9
Right front
±1.1
Right back
±1.1
Center
±0.3
Center of Upper shelf
±0.3
Center of Lower shelf
-
Average
Note:This data does not represent a guarantee of product performance.
<Amount of power consumption>
Amount of power consumption when driving at cycle
(SV=-70℃)
AT23℃
AT30℃
Note:This data does not represent a guarantee
-57-
Ambient temperature 23℃ 
230V,50Hz
Middle of
Maximum
Minimum
Differential Maximum
cycle
-67.3
-70.4
-68.9
±1.6
-67.2
-67.5
-71.1
-69.3
±1.8
-67.5
-67.4
-70.6
-69.0
±1.6
-67.5
-67.4
-70.6
-69.0
±1.6
-67.5
-67.3
-70.6
-69.0
±1.7
-67.3
-69.3
-71.0
-70.2
±0.9
-69.3
-70.0
-71.7
-70.9
±0.9
-69.9
-69.2
-70.8
-70.0
±0.8
-69.2
-70.0
-71.9
-71.0
±1.0
-69.9
-70.1
-71.1
-70.6
±0.5
-70.2
-67.4
-69.6
-68.5
±1.1
-67.4
-68.0
-70.3
-69.2
±1.2
-68.0
-67.7
-69.8
-68.8
±1.1
-67.6
-68.1
-70.6
-69.4
±1.3
-68.1
-67.6
-69.8
-68.7
±1.1
-67.6
-69.5
-70.6
-70.1
±0.6
-69.5
-69.5
-70.5
-70.0
±0.5
-69.5
-
-
-69.5
-
-
Ambient temperature 30℃ 
230V,50Hz
Middle of
Maximum
Minimum
Differential Maximum
cycle
-67.0
-70.1
-68.6
±1.6
-67.0
-67.3
-70.7
-69.0
±1.7
-67.3
-67.4
-70.4
-68.9
±1.5
-67.3
-67.3
-70.4
-68.9
±1.6
-67.3
-67.1
-70.3
-68.7
±1.6
-67.1
-69.4
-71.0
-70.2
±0.8
-69.4
-70.1
-71.7
-70.9
±0.8
-70.1
-69.4
-70.7
-70.1
±0.6
-69.4
-70.0
-71.8
-70.9
±0.9
-70.1
-70.2
-71.0
-70.6
±0.4
-70.2
-67.4
-69.4
-68.4
±1.0
-67.4
-68.0
-70.2
-69.1
±1.1
-68.0
-67.8
-69.8
-68.8
±1.0
-67.8
-68.2
-70.7
-69.5
±1.3
-68.0
-67.7
-69.9
-68.8
±1.1
-67.7
-69.4
-70.4
-69.9
±0.5
-69.4
-69.7
-70.7
-70.2
±0.5
-69.7
-
-
-69.5
-
-
Unit:kWh/day
230V,50Hz
220V,60Hz
5.00
5.10
6.20
6.30
of product performance.
Unit:℃
220V,60Hz
Middle of
Minimum
Differential
cycle
-70.5
-68.9
±1.7
-71.1
-69.3
±1.8
-70.6
-69.1
±1.6
-70.6
-69.1
±1.6
-70.6
-69.0
±1.7
-70.9
-70.1
±0.8
-71.8
-70.9
±0.9
-70.8
-70.0
±0.8
-71.9
-70.9
±1.0
-71.1
-70.7
±0.5
-69.5
-68.5
±1.1
-70.2
-69.1
±1.1
-69.6
-68.6
±1.0
-70.5
-69.3
±1.2
-69.6
-68.6
±1.0
-70.5
-70.0
±0.5
-70.5
-70.0
±0.5
-
-69.5
-
Unit:℃
220V,60Hz
Middle of
Minimum
Differential
cycle
-70.1
-68.6
±1.6
-70.7
-69.0
±1.7
-70.5
-68.9
±1.6
-70.4
-68.9
±1.6
-70.3
-68.7
±1.6
-71.0
-70.2
±0.8
-71.7
-70.9
±0.8
-70.8
-70.1
±0.7
-71.8
-71.0
±0.8
-71.0
-70.6
±0.4
-69.4
-68.4
±1.0
-70.2
-69.1
±1.1
-69.8
-68.8
±1.0
-70.7
-69.4
±1.4
-69.8
-68.8
±1.1
-70.4
-69.9
±0.5
-70.6
-70.2
±0.5
-
-69.5
-

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdf-du502vh

Table of Contents