How to operate Service # 1 and Service # 2 on the service menu is described below. For other operations on the LCD, refer
to the instruction manual.
Term
Runtime
Version UP
Temp. Calibration
System Configuration #1
Status
System Configuration #2
Reset Data
Error Log
Setup Data
Other Setting
CAUTION) Firmware update, compressor, FAN, and cap heater settings affect basic performance.
Be careful when operating.
Function
Display the accumulate time of Main battery, Backup battery FAN A and FAN B, Reset
them.
Update the firmware of main board and touch panel.
Chamber adjustment method.
Setting Capi. Heater and its timer, compressors, Fan A/B .
Monitoring of chamber temp. , ambient temp, comp. evaporator and fan.
Turning On/Off of AIP heater, and Backup. Setting overload alarm delay..
Initialization of data, and flash memory. Setting model code.
Output of error log data.
Input and output the data.
Turning On/Off Screen Capture, Demo Mode, Self-diagnosis Function and FAN Alarm.