Panasonic MDF-DU702VH Service Manual page 71

Ultra-low temperature freezer
Table of Contents

Advertisement

Remove the sensor outlet cover on the back and the temperature
control sensor cover inside the chamber, and remove the silicone
in the backup sensor insertion hole.
Insert the gas thermo into the backup sensor insertion hole from
which silicon has been removed and fix the gas thermo using the
nylon clip attached to the sensor mounting plate.
Fill the gap between sensor insertion hole and gas thermo with
silicon.
Reinstall the temperature control sensor cover and the sensor
outlet cover.
Attach the nylon clips SL - 2 N and SL - 6
N to the holes for those on the back and
secure the gas thermo piping with them.
Insert a notch in the cap A that you
removed, insert a gas thermo piping into it,
and push A into the hole in the base so as
to close the hole.
Remove the seal on the back side and take out the stored
harness PT_AIP_UB_VALVE_ 2.
censor cover
Fig.10-1
thermal control sensor cover
Fig.10-3
nylon crip
SL-2N
nylon crip
SL-6N
solenoid valve power
Fig.12-3
-68-
insert hole for the censor
of back-up kit
installation
Fig.10-4
seal with
silicon
Fig..11-2
make an incision the
cap, insert it into the
pipe, and push the cap
into the hole
Fig.11-1
harnesses
PT_AIP_UB_VALVE_2
Fig.12-1
Fig.12-2
insert hole for gas tube
Fig.10-2
Cation:
direction