2. Checking the Probe Serial No. when Making Settings

Period for which the MR key is
pressed
0
2
Bleep
Bleep
Target color
The probe serial
is displayed.
no. used to mak-
ing settings is
displayed.
<When xyLv, T∆ ∆ ∆ ∆ ∆ uvLv, u'v'Lv or XYZ mode is selected>
The serial number of the probe used when user calibration is performed or when target color is set will be dis-
played.
When the optional 4-Probe Expansion Board CA-B04 is used
The probe serial no. of the probe connector used for the current selected memory channel will be displayed.
$%' $
$%' $
By default (factory setting), "00000000" is set for both 1 and 2 .
• When "00000000" is set for 1 : If measurement is performed with this memory channel, Konica
• When "00000000" is set for 2 : x=0.3127, y=0.3293 and Lv=40.00 (cd/m
<When an analyzer measurement mode is selected>
The probe serial no. that was used to input the analyzer mode RGB emission characteristic or set the target color
for the currently selected memory channel will be displayed.
When the optional 4-Probe Expansion Board CA-B04 is used
The probe serial no. of the probe connector used for the current selected memory channel will be displayed.
$%' $
$%' $
By default (factory setting), "00000000" is set for both 1 and 2 .
• When "00000000" is set for 1 : The RGB emission characteristic for the display's analyzer mode has
* The serial no. of the currently used measuring probe can be viewed in the PROBE selection screen, that can be opened by pressing the
then
keys.
(If the 4-Probe Expansion Board CA-B04 is used, the probe no. will switch from one to another each time the
details, refer to page 43.)
ELECT : PROBE
P1 35881112
To check the probe serial no. when making settings, press the MR key for
two to four seconds (a bleep will sound after two seconds have elapsed)
and check it in the LCD display section.
4
(sec)
Bleep
The unit of
luminance
i s
d i s -
played.
1 Probe serial no. used when user calibration was performed
2 Probe serial no. used when the target color was set
Minolta's calibration standard will be used for the measurement. (Same
as when measurement is performed with the memory channel CH00.)
target color.
1 Probe serial no. used to input the analyzer mode RGB emission characteristic
2 Probe serial no. used when the target color was set
not been input.
73
2
) are set as the values of the
key is pressed. For