Konica Minolta CA-100Plus Instruction Manual page 95

Crt color analyzer
Hide thumbs Also See for CA-100Plus:
Table of Contents

Advertisement

Command
Function
E
ENTER
K
Recalls target color
data from memory
Z
Requests target infor-
m a t i o n r e g a r d i n g
present settings
F
Sets remote-control
mode off or on
Input format
"E"
"K memory channel number "
Acceptable range: 00 to 99
"Z"
"F0": Remote-control mode off
"F1": Remote-control mode on
93
Further information
• Sets present displayed data as tar-
get color data in xyLv or T∆uvLv
display modes (or in analyzer mode
if calibration mode is off).
• Stores previously set RBGW val-
ues in memory when calibration
mode is switched on (if RBGW val-
ues have all been set correctly).
• Corresponds to pressing
• Although
would be pressed to
complete input of data if the operation
were performed using keys, pressing
or inputting the corresponding
command "E" is not necessary. If "E"
is input, the present measured data will
be stored as the target color data.
• This command will not be accepted
unless "I" has been input and zero
calibration has been completed.
• Target color data is output for all
probe numbers selected for data
output.
• For format of output data, see page 95.
• The data output as a result of this
command must be accepted before
the next command is input.
• The following status information
will be output:
a. Display mode
b. SYNC mode
c. Memory channel number
d. ID name
e. Analog display range
f. Luminance units
g. Probe number for which data is
displayed
h. Probe number (or numbers) for
which data is output
i. Calibration mode on or off
j. RBGW values set or not
• For format of output data, see page 96.
• The data output as a result of this
command must be accepted before
the next command is input.
• If remote-control mode is set to off,
the only acceptable input via the RS-
232C terminal is the command "F1".
key.