Turning the Power ON ( | )/OFF (●)
1. Turning the Power ON ( | )/OFF (●)
Before setting the POWER switch to ON ( | ), prepare the following.
1
.
Connect a measuring probe to the probe connector [P1]. (Page 26)
• To synchronize measurement with the ... 1 Input the vertical synchronizing signal that is used for the display.
display's vertical synchronizing signal
(EXT is selected as the SYNC mode)
• To perform measurement ....................... 1 Install the 4-Probe Expansion Board CA-B04 (option) in the
with two or more measuring probes
• To communicate with the PC ................. 1 Connect the instrument's RS-232C connector to the PC. (Page 84)
via RS-232C
• To communicate with the PC via USB ... 1 Connect the instrument's USB connector to the PC. (Page 86)
2
. Connect the AC power cord to an AC outlet. (Page 28)
[Turning the Power ON ( | )]
Set the POWER switch to ON ( | ).
If the instrument is connected to external equipment,
set the instrument's POWER switch to ON ( | ) first,
then turn ON ( | ) the power to the external equipment.
[Turning the Power OFF (●)]
If the instrument is connected to external equipment, turn OFF (●) the power to the external equipment
first, then set the instrument's POWER switch to OFF (●).
<Error Messages in LCD Display Section>
● "SET MAIN PROBE" (After the POWER switch is set to ON ( | ))
• Cause 1 : The measuring probe is not connected to the probe con-
nector [P1] properly.
• Action 1: Set the POWER switch to OFF (●), then connect the measuring probe to the probe connector
[P1] properly. (Before connecting/disconnecting the measuring probe, make sure that the
POWER switch is set to OFF (●).)
● "PROBE ERROR"
• Cause 1 : A measuring probe was connected or disconnected while
the POWER switch was ON ( | ).
• Action 1: Set the POWER switch to OFF (●) first, connect necessary measuring probes, then set the
POWER switch to ON ( | ). (Before connecting/disconnecting the measuring probe, make sure
that the POWER switch is set to OFF (●).)
(Page 28)
instrument. (Page 27)
2 Connect necessary number of probes to the probe connec-
tors [P2] to [P5]. (Pages 26 and 27)
CA-200 MODE
PROBE [P1]
NO.XXXXXXXX
DARKEN PROBE
PUSH 0-CAL KEY
... For other error messages, refer to page 107.