Calibration History
Go to: Menu > Calibration history
Here you can view an instrument's history of calibrations and calibration checks.
Calibration Settings
Go to Menu > System settings > Advanced settings > Calibration for calibration settings.
Automatic Calibration Check
The instrument is set to perform a calibration check automatically by default.
To change this setting, set Automatic check to Disabled.
Calibration Reminder
The instrument is set to remind you when it is time to calibrate by default.
To change the settings:
1. Enable Service mode.
2. Go to: Calibration.
3. Edit the settings for Calibration reminder and Calibration interval as desired.
Custom Calibration
The instrument is set to use Sound Calibrator Type 4231 by default.
To use a different calibrator and set a custom calibration level:
1. Go to: Calibrator.
2. Select Custom.
3. Select Custom calibrator no. to add the calibrator's serial number.
4. Select Custom level to specify the sound level in dB.
Check the
57
Calibration:
Calibration History