Testing The Assembly - Agilent Technologies 75000 C Series Service Manual

32-channel, 5-amp, form c switch
Hide thumbs Also See for 75000 C Series:
Table of Contents

Advertisement

Testing the
Assembly
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
You can use the tests and checks in Table 4-2 to isolate the problem. See
Figures 3-1, 3-2, and 3-3 in Chapter 3 for locations of mechanical parts.
See the component locator included with this manual for locations of
electrical components.
Table 4-2. Agilent E1463A Tests/Checks
Test/Check
Reference Designator
Heat Damage
- - - - - - - - - -
Switch/Jumper
J101, J201
Settings
SP101
Switch Module
F601, F602
PCA
P1, P2, J1, J2
K0-K31
Checking for Heat Damage
Inspect the assembly for signs of abnormal internally generated heat such as
discolored printed circuit boards or components, damaged insulation, or
evidence of arcing.
Checking Switches/Jumpers
Verify that the logical address switch is set correctly (factory set at 120).
Verify that the interrupt priority jumpers are set correctly (factory set at
level 1).
Checking the Switch Module PCA
Check the following:
Verify that fuses F601 and F602 are good.
Check connectors J1, J2, P1, and P2 for damage.
Perform Test 2-1 in Chapter 2.
Check:
Discolored PC boards
Damaged insulation
Evidence of arcing
IRQ Level setting
LADDR setting
Fuse continuity
Connector contacts
Test 2-1, Closed Channel