Example: Closed
Channel Resistance
Test
NOTE
24 Verification Tests
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This example performs a Closed Channel Resistance Test to measure
Channels 0 through 31 relay contact resistances. If the relay contact
resistance for a channel is >2.0
which channel has failed the test. Use this list in Chapter 4 when
troubleshooting a failing relay.
This test assumes that the module is configured without protection
varistors, fuses, or resistors and that jumpers JM1 through JM8 are
installed on the Agilent E1463A Component Assembly (factory setting).
Refer to the Agilent E1463A User's Manual for details and locations of
user installed protection circuits
10!
RE-SAVE "CLOS_TEST"
20
ASSIGN @Dmm TO 722
30
ASSIGN @Sw TO 70915
40
DISP CHR$(129)
50
DIM Result(1,31),Path$(1)[2]
60
DATA NO,NC
70
READ Path$(*)
80
Cc$="01"
90
Ph$="0"
100
!
110
!Start test
120
!
130
CLEAR SCREEN
140
PRINT "Install Component Assembly and Test Fixture "
150
PRINT
160
PRINT " 1. Turn Mainframe and Agilent 3458A DMM power OFF"
170
PRINT " 2. Connect GPIB Cable between mainframe and DMM"
180
PRINT " 3. Install E1463A Component Assembly into Mainframe"
190
PRINT " 4. Attach Test Fixture to Component Assembly"
200
PRINT " 5. Turn Mainframe and DMM power ON "
210
PRINT " 5. Press Continue when ready to begin testing "
220
PAUSE
230
CLEAR SCREEN
240
!
250
! Start the measurements
260
!
270
FOR J = 0 TO 1
280
OUTPUT @Dmm;"PRESET NORM;FUNC OHMF"
(Continued on Next Page)
the program prints a message indicating
! Card number
! Place holder
Chapter 2