Agilent Technologies 75000 C Series Service Manual page 29

32-channel, 5-amp, form c switch
Hide thumbs Also See for 75000 C Series:
Table of Contents

Advertisement

26 Verification Tests
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
680
IF J<1 THEN
690
IF I <10 THEN
700
OUTPUT @Sw;"OPEN (@"&Cc$&Ph$&VAL$(I)&")"
710
ELSE
720
OUTPUT @Sw;"OPEN (@"&Cc$&VAL$(I)&")"
730
END IF
740
ELSE
750
IF I <10 THEN
760
OUTPUT @Sw;"CLOS (@"&Cc$&Ph$&VAL$(I)&")"
770
ELSE
780
OUTPUT @Sw;"CLOS (@"&Cc$&VAL$(I)&")"
790
END IF
800
END IF
810
OUTPUT @Sw;"*OPC?"
820
ENTER @Sw;Opc
830
IF Opc<>1 THEN GOTO 810
840
IF Result(J,I)<0 THEN Result(J,I)=0
850
IF Result(J,I)>2.0 THEN
860
PRINT "Resistance for Channel ";I;" "&Path$(J)&" contacts
is >2.0 Ohms"
870
END IF
880
NEXT I
890
PRINT "Measurements complete for Channel 00-31 "&Path$(J)&" contacts"
900
IF J<1 THEN
910
DISP "Press Continue for channels 00-31 Normally Closed
measurements"
920
PAUSE
930
END IF
940
NEXT J
950
PRINT "Closed Contact Resistance measurements complete"
960
DISP "Press continue to print measurement results"
970
PAUSE
980
CLEAR SCREEN
990
!
1000
! Print measurement results
1010
!
1020 Format:IMAGE "CH ",DD,3X,DD.DDDD," Ohms",3X,DD.DDDD," Ohms"
1030 PRINT TABXY(1,3),"
1040 PRINT
1050 FOR I=0 TO 31
1060
PRINT USING Format;I,Result(0,I),Result(1,I)
1070 NEXT I
1080 END
NO Contacts
NC Contacts"
Chapter 2