Example: DC Isolation
Test
30 Verification Tests
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This example performs DC Isolation Tests for C to chassis and NO to NC.
10!
RE-SAVE "DC_ISOL"
20
ASSIGN @Dmm TO 722
30
ASSIGN @Sw TO 70915
40
DISP CHR$(129)
50
Cc$="01"
60
DIM Result(2)
70
PRINT "Equipment Connections "
80
PRINT
90
PRINT " 1. Turn Mainframe and Agilent 3458A DMM power OFF"
100
PRINT " 2. Connect GPIB Cable between mainframe and DMM"
110
PRINT " 3. Install E1463A Component Assembly into Mainframe "
120
PRINT " 4. Attach Test Fixture to Component Assembly"
130
PRINT " 5. Turn Mainframe and Agilent 3458A power ON"
140
DISP "Press Continue when ready to begin testing "
150
PAUSE
160
OUTPUT @Dmm;"OHM 1E9"
170
CLEAR SCREEN
180
!
190
! Measure DC isolation (NC to NO)
200
!
210
OUTPUT @Sw;"*RST"
220
OUTPUT @Sw;"*OPC?"
230
ENTER @Sw;Opc
240
IF Opc<>1 THEN GOTO 220
250
PRINT TABXY(1,1),"NC (&C) to NO DC Isolation "
260
PRINT TABXY(1,3),"1. Connect DMM Input HI lead to test fixture NO"
270
PRINT TABXY(1,4),"2. Connect DMM Input LO lead to test fixture C"
280
DISP " Press Continue when connections are complete"
290
PAUSE
300
OUTPUT @Dmm:"TRIG SGL"
310
ENTER @Dmm;Result(0)
320
CLEAR SCREEN
330
PRINT TABXY(1,1),"NC (&C) to Chassis DC Isolation"
340
PRINT TABXY(1,3),"1. Connect DMM Input HI lead to Chassis"
350
PRINT TABXY(1,4),"2. Connect DMM Input LO lead to test fixture C"
360
DISP " Press Continue when connections are complete"
370
PAUSE
380
OUTPUT @Dmm;"TRIG SGL"
390
ENTER @Dmm;Result(1)
! Card number