Channel 0 C to NC
Measurements
22 Verification Tests
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Set DMM to measure 4-wire Ohms
Turn mainframe power ON
Measure Channel 0 C to NO Resistance
2.
Send
*RST
to Module
Send
CLOS (@nn00)
(typically 01)
Trigger the DMM with
Send
OPEN (@nn00)
(typically 01)
Enter the result in Table 2-1 for channel 0 C to NO
Repeat for Channels 1 through 31 C to NO
3.
Repeat step 2 for channels 1 through 31
Use
OPEN (@nncc)
= channel # (omit leading zeroes in
cc
Figure 2-2. Common to NO Test Connections
Make Hardware Connections
1.
Turn mainframe power OFF
to close channel 0, where
TRIG SGL
and note reading
to open channel 0, where
and
, where
CLOS (@nncc)
)
nn
= card #
nn
nn
= card #
= card # and
nn
Chapter 2