Agilent Technologies 75000 C Series Service Manual page 32

32-channel, 5-amp, form c switch
Hide thumbs Also See for 75000 C Series:
Table of Contents

Advertisement

NC (&C) to Chassis
Isolation
NO (&C) to Chassis
Isolation
Chapter 2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Make hardware connections as shown in Figure 2-5
1.
Set DMM to 2-wire ohms, 1 G
2.
Send
*RST
to Module
3.
Trigger the DMM with
4.
Record the DMM reading in Table 2-1 (NC to Chassis)
5.
Figure 2-5. NO and NC to Chassis Isolation
Make hardware connections as shown in Figure 2-5
1.
Set DMM to 2-wire ohms, 1 G
2.
Send
*RST
to Module
3.
Send
4.
CLOS (@100:131)
Trigger the DMM with
5.
Record the DMM reading in Table 2-1 (NO to Chassis)
6.
range
TRIG SGL
range
to Module to close all channels
TRIG SGL
Verification Tests 29