To verify printing functions:
19.
Adjust the printer to the following communication settings:
Input
Data Length
Parity Settings
Baud
20.
Using the Communication Adapter Kit, connect the printer to the
monitor. (Refer to Figure 10-6: Connecting the monitor to Printer/
PC with Communication Adapter Kit)
Figure 10-6: Connecting the monitor to Printer/PC
with Communication Adapter Kit
21.
Turn the printer ON and select on-line mode.
22.
Access the monitor's print functions as described in Table 10-5:
"Accessing Monitor's Print Functions":
Section 10: Performance Verification
Serial
8 bits
None
9600 bps
10-9