Changing Default Alarm Limits - Nellcor NPB-75 Service Manual

Handheld capnograph / pulse oximeter
Table of Contents

Advertisement

4.3.1
Default Alarm Limits
The service technician can change the default alarm limits for either
patient mode (Adult or Neonatal) as described in Table 4-3: "Changing
Default Alarm Limits".
The monitor has the following alarms with adjustable level settings:
• EtCO
high and low levels (mmHg)
2
• FiCO
high level (mmHg)
2
• SpO
high and low levels (%)
2
• Pulse rate high and low levels (bpm)
Warning: Changing default settings will permanently change the
user's start-up value and should be verified with the responsible
clinical personnel.
Table 4-3: Changing Default Alarm Limits
Objective
To access the Default
Alarm Limits menu
after entering last
password digit
To change the patient
mode
To access a new
parameter
To change the value
Action
Response
short press
short press
short press
short press
Section 4: Service Mode
4-7