Flow Calibration Process - Nellcor NPB-75 Service Manual

Handheld capnograph / pulse oximeter
Table of Contents

Advertisement

Section 6: Flow Calibration Check
6.3

FLOW CALIBRATION PROCESS

6-2
To perform the Flow Calibration Process, refer to Table 6-1: "Flow
Calibration".
Objective
To access Flow
Calibration screen
To change the
parameter setting or
value
To move to the next
parameter (and save
the changed setting)
To select Flow Cal.
To start Calibration
Wait until either a Pass or Fail message is displayed.
Table 6-1: Flow Calibration
Action
After accessing
Service Level 2
(after entering the
password)
Simultaneously
long press
and
short press
short press
short press
short press
short press
Response