Accessing Flow Calibration - Nellcor NPB-75 Service Manual

Handheld capnograph / pulse oximeter
Table of Contents

Advertisement

Section 4: Service Mode
4.3.3
Flow Calibration
4-10
The Flow Calibration screen allows the service technician to perform flow
calibration after flow adjustment and/or pump or flow system
replacement, refer to Table 4-5: "Accessing Flow Calibration".
Note:
When changing the pump voltage setting, refer to
Flow Calibration Check
Table 4-5: Accessing Flow Calibration
Objective
To access the Flow
Calibration screen
To change the
parameter setting or
value
To move to the next
parameter (and save
the changed setting)
for change guidelines.
Action
Response
At any screen after
the password
long press
simultaneously
and
short press
short press