10.
Reconnect the FilterLine and verify that the message FilterLine
disappears and the pump is operating.
11.
Disconnect the DS-100A - Durasensor
SpO
Sensor is displayed in the Silent Advisory Message area.
2
12.
Reconnect the DS-100A - Durasensor
SpO
Sensor disappears.
2
13.
Verify that all buttons on the unit work properly. Refer to the
monitor's Quick Guide for their applications. (See Figure 10-5:

Quick Guide.)

Figure 10-5: Quick Guide
14.
Verify the CO
Calibration Check process functions. Follow the
2
procedure as described below. When completed, return to the
Measuring Mode screen.
Section 10: Performance Verification
®
and verify that the message
®
and verify that the message
10-3