Section 10: Performance Verification
10-8
15.
Advance to the Plethysmograph Display screen as described in
Table 10-4: "Plethysmograph Waveform" below.
Table 10-4: Plethysmograph Waveform
To View
CO
waveform
2
(Measuring
Mode)
Plethysmograph
short press (x4)
16.
Connect the DS-100A - Durasensor
value appears on the LED digital display and the Pulse rate
appears on the LCD graphic display.
17.
Disconnect the monitor from the AC power supply. Verify the unit
continues to work from its battery pack and the battery icon
appears on the lower right corner of the display.
18.
Reconnect the monitor to mains line power with the AC adapter.
Verify that the unit functions, and the plug icon appears.
Note: If the battery pack has a low charge level, the battery icon
will appear instead of the plug. The battery icon will fill indicating
the battery charging process.
Action
No action
required.
®
to your finger. Verify the SpO