Parameter Reference
8.3 DIAGNOSTICS (3.)
8.3.10

EVENT DIAGNOSIS (3.6.3.)

Factory setting:
Setting range:
Purpose:
To assign a diagnosis to a particular event via the handheld programmer:
1. Go to ENABLE (3.5.1.) (Page 115) and select Yes to enable Condensed Mode.
2. Go to EVENT INDEX (3.6.1.) (Page 116) and enter the event index number
corresponding to the event.
3. Go to Event Diagnosis (3.6.3) and choose a Diagnosis option from the table above.
To assign a status to a particular event via SIMATIC PDM:
1. Go to Level Meter > Maintenance & Diagnostics > Condensed Status > Enable and select
Yes to enable Condensed Status Mode.
2. Go to Level Meter > Maintenance & Diagnostics > Allocation Level.
3. Select the Diagnosis line for the selected Event, then choose a Diagnosis option from the
associated pull-down menu.
See Condensed mode diagnosis (Page 191) for more detail.
2)
8.3.11

PEAK VALUES (3.7.)

Description:
To view via SIMATIC PDM:
Peak value
3.7.1. Min. Measured Value
parameter
references:
3.7.2. Max. Measured Value
3.7.3. Minimum Output Value - AIFB1
3.7.4. Maximum Output Value - AIFB1
3.7.5 Minimum Output Value - AIFB2
3.7.6. Maximum Output Value - AIFB2
118
Maintenance alarm
Status OK
Maintenance required
Maintenance demanded
Maintenance alarm
Invalid process conditions
Function check or simulation
Allows you to assign one of the diagnostic options listed below to any of the
events listed in EVENT INDEX (3.6.1.) (Page 116). This allows you to tailor
a device response appropriate for your particular process.(Event Diagnosis
affects Condensed Acyclic Diagnostics and Cyclic Extended Diagnostics
Open the menu View – Device Diagnostics, select Device Status, and click on the tab Device
Status. For more details, see Diagnostics (Page 83).
The minimum recorded Sensor value, reported in units
defined in UNIT (2.3.1.) (Page 93).
The maximum recorded Sensor value, reported in units
defined in UNIT (2.3.1.) (Page 93).
The minimum recorded Output Value from the Analog
Input Function Block 1.
The maximum recorded Output Value from the Analog
Input Function Block 1.
The minimum recorded Output Value from the Analog
Input Function Block 2.
The maximum recorded Output Value from the Analog
Input Function Block 2.
Operating Instructions, 06/2018, A5E34648471-AE
).
2)
LR560 (PROFIBUS PA)