The HBP-1100 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
HBP-1100 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the HBP-1100 as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter (W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be determined using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
1RWH $W 0+] DQG 0+] WKH VHSDUDWLRQ GLVWDQFH IRU WKH KLJKHU
frequency range applies
1RWH These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
38
Separation distance according to frequency of transmitter (m)
 N+] WR  0+]
d = 1,2 P
0,12
0,38
1,2
3,8
12
 0+] WR  0+]
d = 1,2 P
0,12
0,38
1,2
3,8
12
 0+] WR  *+]
d = 2,3 P
0,23
0,73
2,3
7,3
23