Omron HBP-1100 Instruction Manual page 38

Professional blood pressure monitor
Hide thumbs Also See for HBP-1100:

Advertisement

(OHFWURPDJQHWLF ,PPXQLW\ ,(&
,PPXQLW\ WHVW
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
U
 5DWHG YROWDJH
T
of test unit
Power frequency
  +]
magnetic field
IEC 61000-4-8
1RWH 8
is the a.c. mains voltage prior to application of the test level.
T
36
,(&
WHVW OHYHO
“ N9 FRQWDFW
“ N9 DLU
“ N9 IRU
power supply
lines
“ N9 IRU
input/output lines
“ N9
differential mode
“ N9
common mode
<5 % U
T
for 0.5 cycle
40 % U
T
for 5 cycles
70 % U
T
for 25 cycles
<5 % U
T
for 5 sec.
3 A/m (r.m.s)
&RPSOLDQFH
OHYHO
“ N9 FRQWDFW
“ N9 DLU
“ N9 IRU
power supply
lines
“ N9 IRU
input/output lines
“ N9
differential mode
“ N9
common mode
<5 % U
T
for 0.5 cycle
40 % U
T
for 5 cycles
70 % U
T
for 25 cycles
<5 % U
T
for 5 sec.
3 A/m (r.m.s)
(OHFWURPDJQHWLF HQYL-
URQPHQW  JXLGDQFH
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
HBP-1100 requires
continued operation
during power mains
interruptions, it is
recommended that the
HBP-1100 be powered
from an uninterruptible
power supply or
batteries.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.