Demo 1: Voltage-Controlled Oscillator (Vco) Frequency Step Response And Settling; Start; Overview - HP E1725C Getting Started Manual

Time interval analyzers with the hp e1742a/43a/44a/46a/48a analysis software
Hide thumbs Also See for E1725C:
Table of Contents

Advertisement

Chapter 3 Demo Waveforms
Demo 1: Voltage-Controlled Oscillator (VCO) Frequency Step Response
and Settling
Demo 1: Voltage-Controlled Oscillator (VCO)
Frequency Step Response and Settling
This demonstration shows a VCO whose output is stepped from 9 MHz to
10 MHz. The raw data, or "time stamps" were previously recorded and
subsequently saved to a file, along with the measurement setup which
processes the time stamps into frequency deviation measurements.
The following will be demonstrated:
Recalling data files from disk
Using the cursors and markers
Zooming and scaling
Frequency deviation measurements
Settling time measurement
Frequency overshoot measurement

Start

1
Select File>Load Setup... .
2
Open the demo sub directory (c:\hpe174Xa\demo) and click on the file
"vco.stp."
3
Click the OK button.

Overview

The VCO under test has a voltage step applied to the input control voltage
causing the frequency output to step from 9 MHz to 10 MHz. A trigger
signal occurs approximately 116 µsec prior to the voltage step. This
trigger signal is connected to the "Trigger" input of the HP E1740A.
The VCO output signal is connected to input 1 of the HP E1740A, and is
1 volt pk-pk, centered at 0 volts (no DC offset).
Getting Started Guide
3-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1740a

Table of Contents