Optimizing the Fluorescence Detector

Finding the Best Signal Amplification

Increasing the PMTGAIN increases the signal and the noise. Up to a certain
factor the increase in signal is higher than the increase in noise.
The step from gain to gain is equal to a factor of 2 (which is the same as on the
HP 1046A FLD).
In
from the Agilent Technologies isocratic sample which was diluted 1000
times). With increasing PMTGAIN there was an improvement in signal-to-noise
up to 10. Above 10 the noise increased proportionately to the signal with no
improvement in signal-to-noise.
Figure 27
The reason for this is the fact, that quantification of baselines (especially at
low background levels) is not sufficient for statistically working filter
methods. For the best gain, check your solvent under flow conditions with the
auto-gain function. Do not use higher values than proposed by the system, if
not necessary, because of excessive high fluorescence signals.
Use the PMT test to automatically determine the setting.
66
Figure 27
the PMTGAIN was gradually raised from 4 up to 11 (the peak is
Finding Best PMTGAIN for Biphenyl
1100 Series FD Reference Manual
PMT
11
10
9
8
6
4