Introduction to the Fluorescence Detector

Using the EMF Counters

The user-selectable EMF limits for the EMF counters enable the early
maintenance feedback to be adapted to specific user requirements. The useful
lamp burn time is dependent on the requirements for the analysis (high or low
sensitivity analysis, wavelength etc.), therefore, the definition of the maximum
limits need to be determined based on the specific operating conditions of the
instrument.
Setting the EMF Limits
The setting of the EMF limits must be optimized over one or two maintenance
cycles. Initially, no EMF limit should be set. When instrument performance
indicates maintenance is necessary, take note of the values displayed by lamp
counters. Enter these values (or values slightly less than the displayed values)
as EMF limits, and then reset the EMF counters to zero. The next time the EMF
counters exceed the new EMF limits, the EMF flag will be displayed, providing
a reminder that maintenance needs to be scheduled.