AC Current (Non-Sine Wave) Specifications (cont)
Square Wave
Range
p-p
0.01 Hz to 10 Hz
0.047 mA to
10 Hz to 45 Hz
0.65999 mA [1]
45 Hz to 1 kHz
1 kHz to 10 kHz
0.01 Hz to 10 Hz
0.66 mA to
10 Hz to 45 Hz
6.59999 mA [1]
45 Hz to 1 kHz
1 kHz to 10 kHz
0.01 Hz to 10 Hz
6.6 mA to
10 Hz to 45 Hz
65.9999 mA [1]
45 Hz to 1 kHz
1 kHz to 10 kHz
0.01 Hz to 10 Hz
66 mA to
10 Hz to 45 Hz
659.999 mA [1]
45 Hz to 1 kHz
1 kHz to 10 kHz
0.66 A to
10 Hz to 45 Hz
5.99999 A [2]
45 Hz to 1 kHz
1 kHz to 10 kHz
6 A to 41 A [2]
45 Hz to 500 Hz
500 Hz to 1 kHz
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
1-Year Absolute
± (% of output + % of range)
Frequency
Uncertainty,
tcal ± 5 °C,
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
10 + 2
5.0 + 0.5
0.25 +0.5
0.25 + 0.25
10 + 2
5.0 + 0.5
0.25 + 0.5
0.25 + 0.25
10 + 2
5.0 + 0.5
0.25 + 0.5
0.25 + 0.5
10 + 2
0.5 + 1.0
0.5 + 0.5
10 + 2
0.5 + 0.5
1.0 + 1.0
Multi-Function Calibrator
Specifications
Max
Current
Resolution
Two digits
Six digits
Two digits
Six digits
Two digits
Six digits
Two digits
Six digits
31