Connecting The Cylinder To The Compressor - Invacare HomeFill Oxygen System User Manual

Table of Contents

Advertisement

4.3 Connecting the Cylinder to the Compressor

WARNING!
– NEVER use tools of any kind to
connect/disconnect the cylinder and the
compressor. Otherwise, severe injury and/or
damage may occur.
WARNING!
– DO NOT drop oxygen cylinders. Use two hands
when handling/transporting oxygen cylinders.
Otherwise, injury or damage may occur.
– DO NOT use a liquid leak detector to test for
leaks.
1100873-M
Fillport covers not shown for clarity.
1. Examine the cylinder gauge. If the cylinder pressure is
less than 1500 psig, proceed to STEP 2.
For location of the cylinder gauge, refer to the
cylinder's manual.
If the cylinder pressure is greater than 1500
psig, DO NOT attempt to top off this cylinder.
It may not fill.
2. Set the regulator flow dial on the cylinder A to Off.
Refer to the cylinder's manual.
Usage
19

Advertisement

Table of Contents
loading

Table of Contents