Invacare® HomeFill® Oxygen System
Indicator Light Color
FULL - GREEN
ATTENTION - RED (Audible Alarm)
28
Compressor Status
Compressor is On. Cylinder is finished
filling (full) -remove cylinder (FULL light
is On in >10 mins.)
or
Cylinder is not attached properly - check
connection (FULL light is On in <10
mins.).
Compressor is On. Cylinder NOT filling.
Causes:
Full cylinder attached when unit turned
On - replace with empty cylinder
or
Internal failure - contact dealer.
Effect
Cylinder filling stops.
Cannot fill another cylinder until
compressor is turned Off and On again.
O
flowing to user if concentrator is On.
2
Compressor does not begin to fill or
stops filling.
Unit must be turned Off and On before
filling can occur.
O
flowing to user if concentrator is On.
2